Machine Model: DIANOVA X1
Pump Systems: Blood Pump, Heparin Pump
Clinical Applications: Hemodialysis, Hemoperfusion, Ultrafiltration
Comprehensive Monitoring
ONLINE KT/V MONITORING: ensures proper dialysis dose is delivered in real time.
BLOOD PRESSURE MONITORING: automated and non-invasive system to control blood pressure.
BLOOD VOLUME MONITORING: allows ultrafiltration monitoring according to blood volume variations.
BLOOD TEMPERATURE MONITORING: controls body temperature to ensure hemodynamic balance.
NETWORK MONITORING: interface to connect to renal unit software for centralized data management; USB port also available as alternative to download data.
Advanced Support Features
Flat, rotating 15-inch touch screen for increased readability and convenience.
Ingenious, user-friendly software with high degree of monitoring and personalization.
Bicarbonate cartridge connection to optimize hygiene, cleanliness and safety.
Automated, online priming without the use of serum nor personnel intervention.
Heparin pump designed for multiple syringe sizes with automated syringe size detection, self-test alarm system and heparin profiling.
VP air detector with dual ultrasonic and optical monitoring system for precise bubble catching, safe blood return and continuous micro-bubble detection.
Endotoxin filtration system: single stage on DIANOVA X1 model for high purity HD and double-stage on DIANOVA X2 model for HDF online.
40 min back-up battery in the event of power outage.
Facilitated maintenance through smart maintenance mode and clear, simplified organization of electric and hydraulic compartments.
Connection to centralized concentrate supply system.
Technical Data
Blood pump and substitution pump
Blood flow rate: 0 to 650 mL/min; Accuracy: ±10%
Substitution flow rate: 0 to 650 mL/min; Accuracy: ±10%
Substitution flow range: 0 to 39 L/h (HDF online)
Heparin pump
Heparin pump flow: 0 to 10 mL/h; Accuracy: ±0.2 mL/h
Heparin pump flow resolution: 0.1 mL/h
Syringe size: 10 mL, 20 mL, 30 mL, 50 mL
Injection rate: 10 to 1800 mL/h
Dialysis fluid and conductivity
Dialysate flow: 0, 100 to 1000 mL/min; Accuracy: < ±5%
Dialysate flow resolution: 1 mL/min
Dialysate temperature: 33 to 40°C; Accuracy: ±0.5°C
Dialysate conductivity: 12.0 to 18.0 mS/cm; Accuracy: ±0.1 mS/cm
Ultrafiltration control
UF rate: 0 to 6000 mL/h
Monitoring system
VP: -700 to 800 mmHg; Accuracy: 5 mmHg
AP: -700 to 800 mmHg; Accuracy: ±5 mmHg
TMP: -700 to 800 mmHg; Accuracy: ±5 mmHg
Blood leakage monitor: > 0.35 mL/min (HCT 32%)
Air detector: ≥ 0.02 mL bubbles or ≥ 0.0003 mL continuous small bubbles
Disinfection and cleaning
Rinse: Water / 37°C / 700 mL/min
Hot Rinse: Water/93°C / 700 mL/min
Decalcification: 20%-30%-50% Citric Acid / 37°C / 700 mL/min
Thermal disinfection: 20%-30%-50% Citric Acid / 93°C / 700 mL/min
Chemical disinfection: 5%-10% Sodium Hypochlorite / room temperature / 700 mL/min
Blood pressure monitoring (Non-Invasive)
Blood pressure range: 0 to 300 mmHg
Resolution: 1 mmHg
Pulse monitoring range: 30 to 220 bpm
Pulse monitoring accuracy: ± 3 bpm
Dimensions and weight
Dimensions: 1460 x 490 x 715 mm
Weight: approx. 85 kg
Water supply
Pressure: 0.5 to 8.0 bar
Temperature: 5 to 30 °C
Power supply
Main voltage: 100 to 240 V AC
Frequency: 50 to 60 Hz
Back-up battery: 40 min
Running environment
Temperature: 10 to 40 °C
Humidity: 75%
Atmospheric pressure: 70 kPa to 106 kPa (700 to 1060 mbar)
Protective grade: IPX1
Heater Power: 1800 W